Smart-01 Intelligent Temperature-Humidity Controller
Smart-01 is a temperature and humidity control system independently developed by our SST. Built on an industrial-grade intelligent PLC architecture, it offers excellent reliability, stability, a high degree of customization, and powerful scalability. In addition, Smart-01 features a user-friendly human–machine interface, supports multiple communication protocols, and enables remote monitoring and data acquisition, making operation highly convenient and efficient. This system is widely applied across our full range of environmental test equipment, including: Temperature and humidity chambers, Combined environmental test chambers, Walk-in environmental chambers, Whole vehicle test systems.

Modern & Intuitive Interface
10.1"/15.6" capacitive touchscreen displays
High resolutions: 1280x800P / 1920x1080P
User-friendly HMI with Windows 10 operating system
Multilingual & Adaptive
Flexible for global use across different user environments
Power Failure Recovery
Configurable restart modes: Stop, Start, or Pause after power is restored
Multi-Level Security
Supports over 20 user accounts with 3 access levels, ensuring system safety and traceability
Intelligent Temperature Control
Default mode: air temperature control
Optional mode: specimen surface temperature control for precise testing
Auto Dew Point Control
Automatically manages dew point to prevent condensation on specimens
Communication & Remote Capabilities
Built-in LAN port supporting Ethernet, Modbus TCP, and 3G/4G communication
Compatible with PC for real-time curve display, data logging, and remote monitoring
Remote control via specialized software and hardware over network
Optional support for RS-485, RS-232, and other common protocols
Data Storage & Program Capacity
Data Logging
USB support for downloading historical curves, parameters, and alarm logs (hot-swappable)
Internal large-capacity storage for up to 12 months of historical data
Default data sampling interval: 1 second
Program Settings
Constant run time: Up to 32,000 minutes (or continuous operation)
Single program capacity: 100 steps per group
Total memory: Up to 100 groups / 10,000 steps
Loop execution: Each command supports up to 32,000 cycles
Linking programs supported for complex test profiles